FAQ within [ HPLC ]
-
(LC Column) General method for washing reversed-phase column
・Flush the solution that elutes the substance adsorbed on the column. Do not precipitate out the salt inside the column. If a salt may have been used, the co... Show Detail
- No:10878
- Date of Issue:2024/09/30 09:32
- Category: LC Columns , LC Columns
-
(LC column) Retention time shifted rapidly when the aqueo...
It is recommended to increase the concentration of organic solvents or use a column that supports 100% aqueous mobile phase. Recommended product: Shim-pack ... Show Detail
- No:10682
- Date of Issue:2024/09/30 09:32
- Category: LC Columns , LC Columns
-
(LC Column) Poor retention time reproducibility in batch ...
->Please check whether the mobile phase is degassed. (Reference:5 Degassing Methods : SHIMADZU (Shimadzu Corporation)) ->Please check whether the pum... Show Detail
- No:10681
- Date of Issue:2024/09/30 09:32
- Category: LC Columns , LC Columns
- Brached FAQ
-
(LC Column) Some peaks have bad shapes.
Repeat the analysis to see if there is any improvement. The column may age and improve. If metal adsorption is suspected, try a filler with less metal ions ... Show Detail
- No:10687
- Date of Issue:2024/09/30 09:31
- Category: LC Columns , LC Columns
- Brached FAQ
-
Nexera 40 Series Operation mode
If you are using a new instrument and the software does not support it, you cancontrol it by setting the following operation modes. Please refer the manual ... Show Detail
- No:9171
- Date of Issue:2023/06/09 15:04
- Category: Common
-
(LC) Setting IE mode in Microsoft Edge (Web control)
NOTE: • To use IE mode in Edge, you need to have Chromium Edge(icon ) version 102.0.1245.30or later. • If the Edge icon is , you cannot use IE mo... Show Detail
- No:8986
- Date of Issue:2022/07/28 09:53
- Last Updated:2023/09/11 14:45
- Category: System Controller
-
(LC) Error Messages (pH monitor)
« Error Messages» 8001: PHM: pH sensorerror. (FATAL) The sensor head isnot functioningproperly. →If the error persists, turn off the instrument a... Show Detail
- No:8533
- Date of Issue:2022/03/14 14:26
- Category: pH Monitor
-
Replace the pH electrode. Click here for "(LC) Replacing the pH Electrode". Rinse the pH electrode. Click here for "(LC) Cleaning the pH Electrode". ... Show Detail
- No:8532
- Date of Issue:2022/03/14 14:07
- Category: pH Monitor
- Brached FAQ
-
(LC) Calibrating the Leak Sensor (pH monitor)
1. In the main window of pHM-40 Utility, click the (LeakSensor Calibration) button. The [Leak Sensor Calibration] window will be displayed. In the... Show Detail
- No:8531
- Date of Issue:2022/03/14 13:52
- Category: pH Monitor
-
(LC) Calibrating the Temperature Sensor (pH monitor)
Calibrate the temperature sensor of this instrument. This instrument uses the temperaturesensor connected to a measuring device to detect the temperature of ... Show Detail
- No:8530
- Date of Issue:2022/03/14 13:48
- Category: pH Monitor