• Font Size
  • S
  • M
  • L

Frequently Referenced FAQ

FAQ within [ HPLC ]

171-180 of Total 943

≪ 18 / 95page ≫
  • Error Code:[0x9050]

    «Main Cause» The buffer memory used for transmitting data from the instrument to the workstation is full. «Countermeasure» Check the network connection stat... Show Detail

  • Error Code:[0x9051]

    «Main Cause» The buffer memory used for transmitting data from the instrument to the workstation is full. «Countermeasure» Check the network connection stat... Show Detail

  • Error Code:[0x9052]

    «Main Cause» The IP address is set to "192.168.200.96" because the instrument failed to automatically obtain an IP address. «Countermeasure» Contact the ne... Show Detail

  • Error Code:[0x9053]

    «Main Cause» IP address renewal/refresh failed. «Countermeasure» Contact the network administrator. Set the IP address settings (automatic acquisition/manua... Show Detail

  • Error Code:[0x9074]

    «Main Cause» The system check is timed out. «Countermeasure» Execute system check again. If the error is not resolved, contact a Shimadzu sales or service r... Show Detail

  • Error Code:[0x9075]

    «Main Cause» Purge was executed because the replacement pump pressure reached the [minimum pressure] in the method parameters. «Countermeasure» Inspect for ... Show Detail

  • Error Code:[0x9055]

    «Main Cause» More than 10 instruments are connected to the same group. «Countermeasure» Check the number of instruments connected to the instrument group wh... Show Detail

  • Error Code:[0x9076]

    «Main Cause» The injection volume setting value exceeds the [Max Injection Volume] setting in the system settings. «Countermeasure» Change the injection vol... Show Detail

  • Error Code:[0x9077]

    «Main Cause» A vial number that is different from the sample vial number set in the sequence may have been notified to the instrument. «Countermeasure» Clea... Show Detail

  • Error Code:[0x9078]

    «Main Cause» A rack number that is different from the rack number set in the sequence may have been notified to the instrument. «Countermeasure» Clear the e... Show Detail

171-180 of Total 943

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Technology Inc.