«Main Cause» Failed to analyze the high-speed LC mode. «Countermeasure» Set the gradient control mode to standard mode.If you need the high-speed LC mode, c... Show Detail
- No:4144
- Date of Issue:2020/12/02 16:46
- Category: System Controller
-
«Main Cause» The cell is not fully connected or the flow cell memory is not functioning properly. «Countermeasure» Check the flow cell connection. If that d... Show Detail
- No:4143
- Date of Issue:2020/12/02 16:46
- Category: Detector
-
«Main Cause» A flow cell not compatible with the instrument was connected. «Countermeasure» Connect a flow cell for PDA detectors. Show Detail
- No:4142
- Date of Issue:2020/12/02 16:46
- Category: Detector
-
«Main Cause» The illumination circuit is not functioning properly. «Countermeasure» Replace the D2 lamp. Click here for Replacement method of... Show Detail
- No:4141
- Date of Issue:2020/12/02 16:46
- Last Updated:2020/12/07 11:08
- Category: Detector
-
«Main Cause» The illumination circuit is not functioning properly. «Countermeasure» Replace the D2 lamp. Click here for Replacement method of... Show Detail
- No:4140
- Date of Issue:2020/12/02 16:46
- Last Updated:2020/12/07 11:07
- Category: Detector
-
«Main Cause» The cell temperature could not be controlled to the temperature specified in the method because a cell that does not support temperature-control... Show Detail
- No:4139
- Date of Issue:2020/12/02 16:46
- Category: Detector
-
«Main Cause» Communication failed between the operation panel and the instrument. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sa... Show Detail
- No:4138
- Date of Issue:2020/12/02 16:46
- Category: Detector
-
«Main Cause» Cannot communicate normally between the instrument and workstation. «Countermeasure» Check the communication environment and operating status b... Show Detail
- No:4137
- Date of Issue:2020/12/02 16:46
- Category: Detector
-
«Main Cause» The D2 lamp failed to illuminate. «Countermeasure» Replace the D2 lamp. Click here for Replacement method of Lamp (HPLC) Show Detail
- No:4136
- Date of Issue:2020/12/02 16:46
- Last Updated:2020/12/07 11:07
- Category: Detector
-
«Main Cause» D2 lamp is not connected. «Countermeasure» Connect the D2 lamp. Show Detail
- No:4135
- Date of Issue:2020/12/02 16:46
- Category: Detector