FAQ within [ HPLC ]

151-160 of Total 943

16 / 95page
  • Error Code:[0x528C] FATAL SYRINGE HOME

    «Main Cause» Autosampler: The syringe unit failed to detect the home position. «Countermeasure» Check the syringe plunger and plunger holder for foreign obj... Show Detail

    • No:4146
    • Date of Issue:2020/12/02 16:46
    • Category: Autosampler
  • Error Code:[0x58F5] FATAL INITIALIZE,D2 ERR

    «Main Cause» Initialization was not able to finish normally. Because the lamp cover was opened during 0th order light detection or an error occurred during l... Show Detail

    • No:4145
    • Date of Issue:2020/12/02 16:46
    • Category: Detector
  • Error Code:[0x5C06]

    «Main Cause» Failed to analyze the high-speed LC mode. «Countermeasure» Set the gradient control mode to standard mode.If you need the high-speed LC mode, c... Show Detail

  • Error Code:[0x5AF3]

    «Main Cause» The cell is not fully connected or the flow cell memory is not functioning properly. «Countermeasure» Check the flow cell connection. If that d... Show Detail

    • No:4143
    • Date of Issue:2020/12/02 16:46
    • Category: Detector
  • Error Code:[0x5AF2]

    «Main Cause» A flow cell not compatible with the instrument was connected. «Countermeasure» Connect a flow cell for PDA detectors. Show Detail

    • No:4142
    • Date of Issue:2020/12/02 16:46
    • Category: Detector
  • Error Code:[0x5AF1]

    «Main Cause» The illumination circuit is not functioning properly. «Countermeasure» Replace the D2 lamp. Click here for Replacement method of... Show Detail

    • No:4141
    • Date of Issue:2020/12/02 16:46
    • Last Updated:2020/12/07 11:08
    • Category: Detector
  • Error Code:[0x5AF0]

    «Main Cause» The illumination circuit is not functioning properly. «Countermeasure» Replace the D2 lamp. Click here for Replacement method of... Show Detail

    • No:4140
    • Date of Issue:2020/12/02 16:46
    • Last Updated:2020/12/07 11:07
    • Category: Detector
  • Error Code:[0x5AE9]

    «Main Cause» The cell temperature could not be controlled to the temperature specified in the method because a cell that does not support temperature-control... Show Detail

    • No:4139
    • Date of Issue:2020/12/02 16:46
    • Category: Detector
  • Error Code:[0x5AE8]

    «Main Cause» Communication failed between the operation panel and the instrument. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sa... Show Detail

    • No:4138
    • Date of Issue:2020/12/02 16:46
    • Category: Detector
  • Error Code:[0x5AE7]

    «Main Cause» Cannot communicate normally between the instrument and workstation. «Countermeasure» Check the communication environment and operating status b... Show Detail

    • No:4137
    • Date of Issue:2020/12/02 16:46
    • Category: Detector