Error Code:[0x5296] ERROR FCV SETTING
«Main Cause» FCV port setting is not appropriate. «Countermeasure» Check the setting of FCV ports. Show Detail
- No:4156
- Date of Issue:2020/12/02 16:46
- Category: Autosampler
-
Error Code:[0x5295] ERROR DL.V ERROR
«Main Cause» The DL.V valve connected to the autosampler has detected an error. «Countermeasure» Check the power supply of the column oven where the DL.V va... Show Detail
- No:4155
- Date of Issue:2020/12/02 16:46
- Category: Autosampler
-
Error Code:[0x5294] FATAL H/W ERROR XXXX
«Main Cause» Error in H/W. «Countermeasure» Turn off the instrument and contact a Shimadzu sales or service representative. Show Detail
- No:4154
- Date of Issue:2020/12/02 16:46
- Category: Autosampler
-
Error Code:[0x5293] ERROR FCVx IS LOCAL
«Main Cause» FCV became local mode during analysis. «Countermeasure» Check if the remote LED of the FCV is lit. Show Detail
- No:4153
- Date of Issue:2020/12/02 16:46
- Category: Autosampler
-
Error Code:[0x5292] ERROR FCVx DISCONNECED
«Main Cause» The cause is that communication with the FCV whose [FCV FIX] setting is [1: ON] is cut off. «Countermeasure» Make sure the FCV is turned on. Ch... Show Detail
- No:4152
- Date of Issue:2020/12/02 16:46
- Category: Autosampler
-
Error Code:[0x5291] ERROR ACQUISITION TIME
«Main Cause» Autosampler: Repeated injections did not complete within the data acquisition time. «Countermeasure» Set the injection interval to a time longe... Show Detail
- No:4151
- Date of Issue:2020/12/02 16:46
- Category: Autosampler
-
Error Code:[0x5290] FATAL SYRINGE DATA
«Main Cause» Autosampler: The data stored in the syringe unit is not correct, so the data was initialized. «Countermeasure» Although this does not affect th... Show Detail
- No:4150
- Date of Issue:2020/12/02 16:46
- Category: Autosampler
-
Error Code:[0x528F] ERROR SYRINGE LEAK
«Main Cause» Autosampler: The syringe unit detected leak. «Countermeasure» Wipe off the leaking liquid and check the leaking part. If fluid is leaking from ... Show Detail
- No:4149
- Date of Issue:2020/12/02 16:46
- Category: Autosampler
-
Error Code:[0x528E] FATAL SYRINGE LEAK SEN
«Main Cause» Autosampler: There is an error in the leak sensor of the syringe unit. The leakage sensor may be broken. «Countermeasure» Turn off the instrume... Show Detail
- No:4148
- Date of Issue:2020/12/02 16:46
- Category: Autosampler
-
Error Code:[0x528D] FATAL SYRINGE LPV HOME
«Main Cause» Autosampler: The 3 way valve of the syringe unit failed to detect the home position. «Countermeasure» Turn off the instrument and contact a Shi... Show Detail
- No:4147
- Date of Issue:2020/12/02 16:46
- Category: Autosampler
141-150 of Total 943