  • Error Code:[0x9216]

    «Main Cause» Temperature control is turned off because the sensor in the autosampler dehumidifier unit has detected an abnormal condition. «Countermeasure» ... Show Detail

    • No:3484
    • Date of Issue:2020/12/02 16:10
    • Category: Autosampler
  • Error Code:[0x9217]

    «Main Cause» The backup parameters for the autosampler and column oven are corrupted. «Countermeasure» Press the (power switch) to turn the power OFF and co... Show Detail

  • Error Code:[0x9218]

    «Main Cause» The consumable part count data for the autosampler and column oven is corrupted. «Countermeasure» Press the (power switch) to turn the power OF... Show Detail

  • Error Code:[0x9219]

    «Main Cause» The front panel of the autosampler may not be attached correctly. Alternatively, the sensor that detects the position of the front panel is malf... Show Detail

    • No:3487
    • Date of Issue:2020/12/02 16:10
    • Category: Autosampler
  • Error Code:[0x921A]

    «Main Cause» Temperature control is turned off because a rotation error was detected in the cooling fan of the autosampler. «Countermeasure» Press the (powe... Show Detail

    • No:3488
    • Date of Issue:2020/12/02 16:10
    • Category: Autosampler
  • Error Code:[0x921B]

    «Main Cause» A rotation error was detected in the exhaust heat fan at the top of the rear of the instrument. «Countermeasure» Press the (power switch) to tu... Show Detail

    • No:3489
    • Date of Issue:2020/12/02 16:10
    • Category: Autosampler
  • Error Code:[0x921C]

    «Main Cause» Could not detect the home position of the autosampler high pressure valve. «Countermeasure» Press the (power switch) to turn the power OFF and ... Show Detail

    • No:3490
    • Date of Issue:2020/12/02 16:10
    • Category: Autosampler
  • Error Code:[0x921D]

    «Main Cause» Teaching has not been executed for the autosampler injection port. «Countermeasure» Adjust the position of the needle in relation t... Show Detail

    • No:3491
    • Date of Issue:2020/12/02 16:10
    • Last Updated:2021/01/19 12:52
    • Category: Autosampler
  • Error Code:[0x9400]

    «Main Cause» The column oven door is open when the column oven is above 50ºC. «Countermeasure» Clear the error. The alarm sound and error message disappear.... Show Detail

    • No:3492
    • Date of Issue:2020/12/02 16:10
    • Category: Column Oven
  • Error Code:[0x9402]

    «Main Cause» An error was detected with the flow channel selection valve in the column oven. «Countermeasure» Press the (power switch) to turn the power OFF... Show Detail

    • No:3493
    • Date of Issue:2020/12/02 16:10
    • Category: Column Oven