• Font Size
  • S
  • M
  • L

FAQ within [ HPLC ]

81-90 of Total 929

9 / 93page
  • Error Code:[0x93D8]

    «Main Cause» Temperature control is turned off because the temperature of the optical unit is high. «Countermeasure» Press the (power switch) to turn the po... Show Detail

    • No:3434
    • Date of Issue:2020/12/02 16:10
    • Category: Detector
  • Error Code:[0x93D9]

    «Main Cause» Temperature control is turned off because the temperature sensor of the optical unit is malfunctioning. «Countermeasure» Press the (power switc... Show Detail

    • No:3435
    • Date of Issue:2020/12/02 16:10
    • Category: Detector
  • Error Code:[0x93DA]

    «Main Cause» The instrument is turned off because the ADC module on the board is malfunctioning. The instrument is stopped. «Countermeasure» Press the (pow... Show Detail

    • No:3436
    • Date of Issue:2020/12/02 16:10
    • Category: Detector
  • Error Code:[0x93DB]

    «Main Cause» The optical unit temperature does not reach the set temperature. Temperature control is turned off. «Countermeasure» Contact a Shimadzu sales o... Show Detail

    • No:3437
    • Date of Issue:2020/12/02 16:10
    • Category: Detector
  • Error Code:[0x93DC]

    «Main Cause» The heater of the spectrmeter is short or there is a fault in the circuit. «Countermeasure» Press the (power switch) to turn the power OFF and ... Show Detail

    • No:3438
    • Date of Issue:2020/12/02 16:10
    • Category: Detector
  • Error Code:[0x93E2]

    «Main Cause» Failed to receive the detector signal. «Countermeasure» Press the (power switch) to turn the power OFF and contact a Shimadzu sales or service ... Show Detail

    • No:3439
    • Date of Issue:2020/12/02 16:10
    • Category: Detector
  • Error Code:[0x93E4]

    «Main Cause» Failed to receive the detector signal. «Countermeasure» Press the (power switch) to turn the power OFF and contact a Shimadzu sales or service ... Show Detail

    • No:3440
    • Date of Issue:2020/12/02 16:10
    • Category: Detector
  • Error Code:[0x93F1]

    «Main Cause» Could not detect the filter home position in the detector. «Countermeasure» Press the (power switch) to turn the power OFF and contact a Shima... Show Detail

    • No:3441
    • Date of Issue:2020/12/02 16:10
    • Category: Detector
  • Error Code:[0x93F2]

    «Main Cause» Could not detect the grating home position in the detector. «Countermeasure» Press the (power switch) to turn the power OFF and contact a Shim... Show Detail

    • No:3442
    • Date of Issue:2020/12/02 16:10
    • Category: Detector
  • Error Code:[0x93F4]

    «Main Cause» Results of wavelength accuracy check at instrument startup differ by more than 1nm. There may be bubbles in the flow cell. «Countermeasure» Per... Show Detail

    • No:3443
    • Date of Issue:2020/12/02 16:10
    • Category: Detector

81-90 of Total 929

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.