• Font Size
  • S
  • M
  • L

FAQ within [ HPLC ]

781-790 of Total 929

79 / 93page
  • Error Code:[0x9020]

    «Main Cause» An undefined fatal error has occurred. «Countermeasure» Check whether an error is displayed on the external detector, and follow the steps in... Show Detail

  • Error Code:[0x9071]

    «Main Cause» Acquisition was started when a detector that differs from the detector set in the simple parameters of the acquisition main screen was connected... Show Detail

  • Error Code:[0x9010]

    «Main Cause» Communication is no longer possible with the external detector connected to the optical cable connector on the instrument. «Countermeasure» Che... Show Detail

  • Error Code:[0x9070]

    «Main Cause» An error signal was received from the external device connected to the external input terminal. «Countermeasure» Check in accordance with the h... Show Detail

  • Error Code:[0x9006]

    «Main Cause» Instrument information (system name, group name) was initialized because an error was found in the backup parameters. «Countermeasure» Set the... Show Detail

  • Error Code:[0x9004]

    «Main Cause» The method and sequence are returned to the same condition as when the instrument was last started because the backup parameters are corrupted. ... Show Detail

  • Error Code:[0x9003]

    «Main Cause» The backup parameters of the controller are corrupted. The method file and sequence file were initialized. «Countermeasure» Set the method and... Show Detail

  • Error Code:[0x940A]

    «Main Cause» The room temperature sensor of the instrument has detected an abnormal condition. «Countermeasure» Press the (power switch) to turn the power O... Show Detail

    • No:3495
    • Date of Issue:2020/12/02 16:10
    • Category: Column Oven
  • Error Code:[0x9403]

    «Main Cause» The gas sensor of the column oven has detected an abnormal condition. «Countermeasure» Press the (power switch) to turn the power OFF and conta... Show Detail

    • No:3494
    • Date of Issue:2020/12/02 16:10
    • Category: Column Oven
  • Error Code:[0x9402]

    «Main Cause» An error was detected with the flow channel selection valve in the column oven. «Countermeasure» Press the (power switch) to turn the power OFF... Show Detail

    • No:3493
    • Date of Issue:2020/12/02 16:10
    • Category: Column Oven

781-790 of Total 929

This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.