FAQ - Frequently Asked Questions
- Products Search > HPLC
Keyword Search
HPLC
Frequently Referenced FAQ
FAQ within [ HPLC ]
-
«Main Cause» A flow cell not compatible with the instrument was connected. «Countermeasure» Connect a flow cell for PDA detectors. Show Detail
- No:4142
- Date of Issue:2020/12/02 16:46
- Category: Detector
-
«Main Cause» The illumination circuit is not functioning properly. «Countermeasure» Replace the D2 lamp. Click here for Replacement method of... Show Detail
- No:4141
- Date of Issue:2020/12/02 16:46
- Last Updated:2020/12/07 11:08
- Category: Detector
-
«Main Cause» The illumination circuit is not functioning properly. «Countermeasure» Replace the D2 lamp. Click here for Replacement method of... Show Detail
- No:4140
- Date of Issue:2020/12/02 16:46
- Last Updated:2020/12/07 11:07
- Category: Detector
-
«Main Cause» The cell temperature could not be controlled to the temperature specified in the method because a cell that does not support temperature-control... Show Detail
- No:4139
- Date of Issue:2020/12/02 16:46
- Category: Detector
-
«Main Cause» Communication failed between the operation panel and the instrument. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sa... Show Detail
- No:4138
- Date of Issue:2020/12/02 16:46
- Category: Detector
-
«Main Cause» Cannot communicate normally between the instrument and workstation. «Countermeasure» Check the communication environment and operating status b... Show Detail
- No:4137
- Date of Issue:2020/12/02 16:46
- Category: Detector
-
«Main Cause» The D2 lamp failed to illuminate. «Countermeasure» Replace the D2 lamp. Click here for Replacement method of Lamp (HPLC) Show Detail
- No:4136
- Date of Issue:2020/12/02 16:46
- Last Updated:2020/12/07 11:07
- Category: Detector
-
«Main Cause» D2 lamp is not connected. «Countermeasure» Connect the D2 lamp. Show Detail
- No:4135
- Date of Issue:2020/12/02 16:46
- Category: Detector
-
«Main Cause» The illumination circuit is not functioning properly. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service ... Show Detail
- No:4134
- Date of Issue:2020/12/02 16:46
- Category: Detector
-
«Main Cause» Detector data processing did not function properly. «Countermeasure» Switch the instrument power OFF and contact a Shimadzu sales or service re... Show Detail
- No:4133
- Date of Issue:2020/12/02 16:46
- Category: Detector
141-150 of Total 929