FAQ - Frequently Asked Questions
- Products Search > HPLC
Keyword Search
HPLC
Frequently Referenced FAQ
FAQ within [ HPLC ]
-
Error Code:[0x5292] ERROR FCVx DISCONNECED
«Main Cause» The cause is that communication with the FCV whose [FCV FIX] setting is [1: ON] is cut off. «Countermeasure» Make sure the FCV is turned on. Ch... Show Detail
- No:4152
- Date of Issue:2020/12/02 16:46
- Category: Autosampler
-
Error Code:[0x5291] ERROR ACQUISITION TIME
«Main Cause» Autosampler: Repeated injections did not complete within the data acquisition time. «Countermeasure» Set the injection interval to a time longe... Show Detail
- No:4151
- Date of Issue:2020/12/02 16:46
- Category: Autosampler
-
Error Code:[0x5290] FATAL SYRINGE DATA
«Main Cause» Autosampler: The data stored in the syringe unit is not correct, so the data was initialized. «Countermeasure» Although this does not affect th... Show Detail
- No:4150
- Date of Issue:2020/12/02 16:46
- Category: Autosampler
-
Error Code:[0x528F] ERROR SYRINGE LEAK
«Main Cause» Autosampler: The syringe unit detected leak. «Countermeasure» Wipe off the leaking liquid and check the leaking part. If fluid is leaking from ... Show Detail
- No:4149
- Date of Issue:2020/12/02 16:46
- Category: Autosampler
-
Error Code:[0x528E] FATAL SYRINGE LEAK SEN
«Main Cause» Autosampler: There is an error in the leak sensor of the syringe unit. The leakage sensor may be broken. «Countermeasure» Turn off the instrume... Show Detail
- No:4148
- Date of Issue:2020/12/02 16:46
- Category: Autosampler
-
Error Code:[0x528D] FATAL SYRINGE LPV HOME
«Main Cause» Autosampler: The 3 way valve of the syringe unit failed to detect the home position. «Countermeasure» Turn off the instrument and contact a Shi... Show Detail
- No:4147
- Date of Issue:2020/12/02 16:46
- Category: Autosampler
-
Error Code:[0x528C] FATAL SYRINGE HOME
«Main Cause» Autosampler: The syringe unit failed to detect the home position. «Countermeasure» Check the syringe plunger and plunger holder for foreign obj... Show Detail
- No:4146
- Date of Issue:2020/12/02 16:46
- Category: Autosampler
-
Error Code:[0x58F5] FATAL INITIALIZE,D2 ERR
«Main Cause» Initialization was not able to finish normally. Because the lamp cover was opened during 0th order light detection or an error occurred during l... Show Detail
- No:4145
- Date of Issue:2020/12/02 16:46
- Category: Detector
-
«Main Cause» Failed to analyze the high-speed LC mode. «Countermeasure» Set the gradient control mode to standard mode.If you need the high-speed LC mode, c... Show Detail
- No:4144
- Date of Issue:2020/12/02 16:46
- Category: System Controller
-
«Main Cause» The cell is not fully connected or the flow cell memory is not functioning properly. «Countermeasure» Check the flow cell connection. If that d... Show Detail
- No:4143
- Date of Issue:2020/12/02 16:46
- Category: Detector
131-140 of Total 929