Cause1: The external alarm is working.
Action1: Reset the external device to cancel the alarm output.
Cause2: The polarity of the external input is different.
Action2: If the alarm cannot be cleared even after resetting the external device, restart the controller while holding down the [CLEAR] and [CANCEL] keys of the controller. Initialize the external input settings (set to normal open).
*The digital input settings are also initialized (OFF setting) at the same time.
Cause3: Alarm input is set for digital input.
Action3: Reset the external device or change the digital input settings.
If the alarm cannot be cleared even after resetting the external device, restart the controller while holding down the [CLEAR] and [CANCEL] keys of the controller. Initialize the digital input settings (set to OFF).
*The external input settings are also initialized (normal open) at the same time.