Error Code:[0x0008]
Back
No : 3820
Date of Issue : 2020/12/02 16:43
Print
Error Code:[0x0008]
Error Message: The external input (Alarm Input terminal) of the system controller has an Alarm Input from the external device. Follow the instruction of the external device.
Category :
Products Search
>
HPLC
>
System Controller
Answer
«Main Cause»
An error signal was input from an external device connected to the system controller external input/output terminal ([IN] terminal).
«Countermeasure»
Check the connected external devices and eliminate the cause of the error.