• No : 3389
  • Date of Issue : 2020/12/02 15:47
  • Print

Error Code:[0x5623] FATAL INSIDE FAN

Error Message: Chemical reactor: The temperature control was turned off because of the detection of a rotation error in the cabinet. Remove anything that interferes with the rotation of the fans and restart the instrument.
Category : 

Answer

«Main Cause»
The fan inside the compartment may have stopped.
 
«Countermeasure»
If tubing or other foreign matter is obstructing fan rotation, switch the instrument power OFF and remove the obstruction. If the foreign matter cannot be removed or if the error is not resolved, contact a Shimadzu sales or service representative.