FAQ within [ LC-MS ]
-
Applicable Models:LCMS-8045(RX)/8050(RX)/8060(NX/RX)/8065XE FAQ Title Maintenance Items Video Title Maintenance of the ESI Unit ☞Replacing the... Show Detail
- No:12424
- Date of Issue:2026/07/30 14:04
- Last Updated:2026/08/17 11:14
- Category: Maintenance
-
1. Open the front cover. 2. Check the level of the standard sample. WARNING:Do NOT open the standard sample bottle lid while an MS check is o... Show Detail
- No:8849
- Date of Issue:2022/06/24 09:31
- Category: Setting
- Brached FAQ
-
(LCMS) Removing / Mounting the Ionization Interface
1. Turn the LC solvent delivery pump OFF. 2. Check that an MS check is not running. 3. Open the front cover. 4. Remove the tubing c... Show Detail
- No:8826
- Date of Issue:2022/06/21 11:05
- Category: Maintenance
- Brached FAQ
-
(LCMS) Cleaning the Cooling Fan
1. Remove the dust from the fan filters on the instrument with, for example, avacuum cleaner. Clean the fan filters about every six months. NOTE:W... Show Detail
- No:8823
- Date of Issue:2022/06/20 15:53
- Category: Maintenance
-
(LCMS) Viewing Detailed Actions or Results for System Che...
This content is for LCMS-2050. The following describes how to view detailed actions or results from using the [System Check]-[MS Check] function, pa... Show Detail
- No:7926
- Date of Issue:2022/01/25 10:51
- Category: Software
-
(LCMS) Changing the System Check Pass/Fail Criteria
System check criteria can be changed by following the procedure below. Open the [System Configuration] sub-window, select the LCMS system in the [Modules ... Show Detail
- No:7928
- Date of Issue:2022/01/25 10:49
- Category: Software
-
(LCMS) Checking the MS Instrument Status
This content is for LCMS-2050. The [System Check]-[MS Check] function can automatically check or determine the status of MS detectors (mass axis shift, FW... Show Detail
- No:7927
- Date of Issue:2022/01/25 10:46
- Last Updated:2022/06/24 12:50
- Category: Software
-
(LCMS) Relationship between Method, Batch, and Data Files
Method files (.lcm, .gcm) contain data acquisition conditions, analysis conditions, calibration curve settings, and other information. Batch files (.lcb, .g... Show Detail
- No:7922
- Date of Issue:2022/01/25 09:05
- Category: Software
-
In the [MS Data Analysis] window, click the [Spectrum View]-[Display] tab, and then select the checkboxes for the event numbers to display. To overlay spect... Show Detail
- No:7920
- Date of Issue:2022/01/25 09:02
- Category: Software
-
(LCMS) Displaying MS Chromatograms
Right-click on the [Chromatogram View] in the [MS Data Analysis] window and then click [Data View Parameters] on the menu (or on the toolbar). Click the [... Show Detail
- No:7918
- Date of Issue:2022/01/25 08:53
- Category: Software