FAQ - Frequently Asked Questions
- Products Search > Elemental Analysis
Keyword Search
Frequently Referenced FAQ
FAQ within [ Elemental Analysis ]
-
(PDA) Replacing the Argon Gas Cylinder
When the displayed value for the regulator (1), whichis attached to the argon gas cylinder (hereinafter, thecylinder), becomes 2 MPa or less, replace the cyl... Show Detail
- No:6525
- Date of Issue:2021/06/28 10:40
- Category: Optical Emission Spectrometer (PDA)
-
(PDA) Replacing the Recalibration Sample
1. Print out the [Analytical Group Information] –[Recalibration] windows for the analytical group tobe changed. 2.Clean the emission table. 3.C... Show Detail
- No:6526
- Date of Issue:2021/06/28 11:01
- Category: Optical Emission Spectrometer (PDA)
- Brached FAQ
-
(PDA) Emergency stop and recovering (PDA-7000)
CAUTION: If any problem should occur on the instrument, stop the instrumentby emergency stop and immediately remove the cause of theproblem. 1. Emerge... Show Detail
- No:6527
- Date of Issue:2021/06/28 11:24
- Last Updated:2021/08/02 15:38
- Category: Optical Emission Spectrometer (PDA)
-
(PDA) Daily Startup / Shutdown
1. Turn ON the power to the printer and monitor. 2. Confirm that the argon gas supply pressure is in the range of140 kPa to 200 kPa. Confirm that the ... Show Detail
- No:6528
- Date of Issue:2021/06/28 12:54
- Category: Optical Emission Spectrometer (PDA)
- Brached FAQ
-
(PDA) Startup after a long period shut down
1. Confirm that the room temperature is in the range of 15 °C to 30 °C. 2. Confirm that the argon gas supply pressure is in the range of140 kPa to 200 kPa. ... Show Detail
- No:6529
- Date of Issue:2021/06/28 14:42
- Category: Optical Emission Spectrometer (PDA)
- Brached FAQ
-
1. Open the [Instrument Status] window. 2. Shut down the software. 3. Shut down Windows and turn OFF the PC and peripheral devices. 4. After tu... Show Detail
- No:6530
- Date of Issue:2021/06/28 14:54
- Category: Optical Emission Spectrometer (PDA)
- Brached FAQ
-
(PDA) The spectrometer inner pressure has problem
« Cause» • The O-ring of thecondenser lens is notcorrectly installed. • The oil level in thevacuum pump is not thespecified one. • The pump oil in th... Show Detail
- No:6532
- Date of Issue:2021/07/06 10:44
- Last Updated:2021/08/02 16:24
- Category: Optical Emission Spectrometer (PDA)
-
(PDA) Temperature of the spectrometer is unstable, or doe...
« Cause » • Abnormal temperature ofthe spectrometer • The fan supplying hot airhas a problem. « Measure» PDA-8000: Confirm that the temperat... Show Detail
- No:6536
- Date of Issue:2021/07/06 10:54
- Category: Optical Emission Spectrometer (PDA)
-
(PDA) Analysis accuracy is poor
« Cause » • The argon gas leaks during analysis. • The electrode is contaminated. « Measure» Set the sample correctly and performreanalysis. ... Show Detail
- No:6539
- Date of Issue:2021/07/06 11:06
- Last Updated:2021/08/02 16:26
- Category: Optical Emission Spectrometer (PDA)
-
(PDA) The buzzer on the temperature indicator sounds
« Cause» • Abnormal temperature ofthe spectrometer. « Measure» PDA-8000: Confirm that the temperature of theanalysis chamber is 15 °C to 30 °C.... Show Detail
- No:6541
- Date of Issue:2021/07/06 11:06
- Category: Optical Emission Spectrometer (PDA)
81-90 of Total 137