FAQ within [ Optical Emission Spectrometer (PDA) ]

1-10 of Total 16

1 / 2page
  • (PDA) Basic Points to be Checked When...

    Sometimes, problems are caused by a simple cause. If a problem happens, check the followings first. • Are all power switches turned ON? Confirmtha... Show Detail

    • No:6615
    • Displayed Date:2021/07/14 13:20
  • (PDA) The buzzer on the temperature i...

    « Cause» • Abnormal temperature ofthe spectrometer. « Measure» PDA-8000: Confirm that the temperature of theanalysis chamber is 15 °C to 30 °C.... Show Detail

    • No:6541
    • Displayed Date:2021/07/06 11:06
  • (PDA) Analysis accuracy is poor

    « Cause » • The argon gas leaks during analysis. • The electrode is contaminated. « Measure» Set the sample correctly and performreanalysis. ... Show Detail

    • No:6539
    • Displayed Date:2021/07/06 11:06
    • Updated DTM:2021/08/02 16:26
  • (PDA) Temperature of the spectrometer...

    « Cause » • Abnormal temperature ofthe spectrometer • The fan supplying hot airhas a problem. « Measure» PDA-8000: Confirm that the temperat... Show Detail

    • No:6536
    • Displayed Date:2021/07/06 10:54
  • (PDA) The spectrometer inner pressure...

    « Cause» • The O-ring of thecondenser lens is notcorrectly installed. • The oil level in thevacuum pump is not thespecified one. • The pump oil in th... Show Detail

    • No:6532
    • Displayed Date:2021/07/06 10:44
    • Updated DTM:2021/08/02 16:24
  • (PDA) Long term shutdown

    1. Open the [Instrument Status] window. 2. Shut down the software. 3. Shut down Windows and turn OFF the PC and peripheral devices. 4. After tu... Show Detail

    • No:6530
    • Displayed Date:2021/06/28 14:54
    • Brached FAQ
  • (PDA) Startup after a long period shu...

    1. Confirm that the room temperature is in the range of 15 °C to 30 °C. 2. Confirm that the argon gas supply pressure is in the range of140 kPa to 200 kPa. ... Show Detail

    • No:6529
    • Displayed Date:2021/06/28 14:42
    • Brached FAQ
  • (PDA) Daily Startup / Shutdown

    1. Turn ON the power to the printer and monitor. 2. Confirm that the argon gas supply pressure is in the range of140 kPa to 200 kPa. Confirm that the ... Show Detail

    • No:6528
    • Displayed Date:2021/06/28 12:54
    • Brached FAQ
  • (PDA) Emergency stop and recovering (...

    CAUTION: If any problem should occur on the instrument, stop the instrumentby emergency stop and immediately remove the cause of theproblem. 1. Emerge... Show Detail

    • No:6527
    • Displayed Date:2021/06/28 11:24
    • Updated DTM:2021/08/02 15:38
  • (PDA) Replacing the Recalibration Sample

    1. Print out the [Analytical Group Information] –[Recalibration] windows for the analytical group tobe changed. 2.Clean the emission table. 3.C... Show Detail

    • No:6526
    • Displayed Date:2021/06/28 11:01
    • Brached FAQ