(EDX) In standard sample analysis, there was a large devi...
Do you perform drift correction? If you are using a calibration curve from a few years ago, it is better to revise the calibration curve. Show Detail
- No:5691
- Date of Issue:2021/04/28 13:41
- Last Updated:2021/05/20 08:54
- Category: Energy Dispersive X-ray Fluorescence Spectroscopy(EDX)
-
(EDX) Is it possible to distinguish PBB and PBDE with EDX?
• With the EDX, compounds cannot be distinguished. Perform screening analysis with respect to total Br content. • If knowing the total Brcontent of a sa... Show Detail
- No:5689
- Date of Issue:2021/04/28 13:40
- Last Updated:2021/05/20 08:52
- Category: Energy Dispersive X-ray Fluorescence Spectroscopy(EDX)
-
(EDX) What is mean of Ka and Kb in quantitative profile
These are spectral line names.Multiple lines are generated by a single element. Indicate:Siegbahn notation FeKa: FeKα FeKb: FeKβ FeLa: FeLα FeLb1: FeL... Show Detail
- No:5687
- Date of Issue:2021/04/28 13:36
- Last Updated:2022/06/24 16:15
- Category: Energy Dispersive X-ray Fluorescence Spectroscopy(EDX)
-
(EDX) Ceramic plate sample analysis in resin film
Analysis cannot proceed as far as the resin on the opposite side. (1) The ceramic is in the middle, so X-ray intensity from the resin on the opposite side w... Show Detail
- No:5685
- Date of Issue:2021/04/28 13:33
- Last Updated:2021/05/20 08:49
- Category: Energy Dispersive X-ray Fluorescence Spectroscopy(EDX)
-
(EDX) Mylar film and "Mg" analysis
If a Mylar film is used, the Mg intensity will be significantly reduced. Either use a polypropylene film or a pressure molded briquette. Show Detail
- No:5683
- Date of Issue:2021/04/28 13:32
- Last Updated:2021/05/20 08:48
- Category: Energy Dispersive X-ray Fluorescence Spectroscopy(EDX)
-
(EDX) Thickness of powder sample
Use a micro X-cell. If there is enough sample to extend at least 5mm from the bottom, then the thickness will be sufficient. If the thickness is less than ... Show Detail
- No:5681
- Date of Issue:2021/04/28 13:30
- Last Updated:2021/05/20 08:47
- Category: Energy Dispersive X-ray Fluorescence Spectroscopy(EDX)
-
(EDX) "Na" analysis in aqueous solution (without He subst...
Use the filter instillation method for the analysis. In the filter instillation method, droplets of a liquid are dripped on to a filter paper. Then they are... Show Detail
- No:5679
- Date of Issue:2021/04/28 13:29
- Last Updated:2021/05/20 08:46
- Category: Energy Dispersive X-ray Fluorescence Spectroscopy(EDX)
-
(EDX) Cl detection in "Easy" analysis
The ClKα line overlaps with the RhLα line originating from the X-ray tube. As a result, trace quantities of Cl cannot be detected with a easy analysis method... Show Detail
- No:5677
- Date of Issue:2021/04/28 13:28
- Last Updated:2021/05/20 08:44
- Category: Energy Dispersive X-ray Fluorescence Spectroscopy(EDX)
-
(EDX) Analysis of "Si" and "P" in "Cu" alloy
(1) Set the atmosphere to vacuumfor this analysis. →This will improve the sensitivity for light elements. (2) Lower the tube voltagefor the analysis. → ... Show Detail
- No:5675
- Date of Issue:2021/04/28 13:26
- Last Updated:2021/05/20 08:42
- Category: Energy Dispersive X-ray Fluorescence Spectroscopy(EDX)
-
(EDX) Film settings and quantitative values
If the sample contains light elements, then there may be problems with the quantitative values. •The energy of the fluorescent X-rays from the light element... Show Detail
- No:5673
- Date of Issue:2021/04/28 13:24
- Last Updated:2021/05/20 08:41
- Category: Energy Dispersive X-ray Fluorescence Spectroscopy(EDX)