FAQ within [ Elemental Analysis ]
-
The analysis depth for fluorescent X-rays changes with the principal components of the sample and the target elements (analysis lines). •The lighter the ele... Show Detail
- No:5710
- Date of Issue:2021/04/28 13:57
- Last Updated:2021/05/20 09:05
- Category: Energy Dispersive X-ray Fluorescence Spectroscopy(EDX)
-
(EDX) Unit of "%" in "Easy" analysis
The unit of the values obtained in easy analysis mode is wt%. Quantitative values can also be calculated in atom%. In quantitative mode, this can be set pr... Show Detail
- No:5708
- Date of Issue:2021/04/28 13:55
- Last Updated:2021/05/20 09:04
- Category: Energy Dispersive X-ray Fluorescence Spectroscopy(EDX)
-
• Strongly magnetic samples cannot be measured. •Measurement is possible if the magnetism at a distance of 7cm from the sample is 20 gauss or less. •If pos... Show Detail
- No:5706
- Date of Issue:2021/04/28 13:54
- Last Updated:2022/09/20 13:58
- Category: Energy Dispersive X-ray Fluorescence Spectroscopy(EDX)
-
(EDX) "Hg" detected by RoHS analysis in resin containing ...
•Hg was detected due to the influence of the BrKαESC peak. →HgLα: 9.99 keV, BrKαESC: 10.16 keV •With the EDX, when high intensity peaks are observed, it i... Show Detail
- No:5704
- Date of Issue:2021/04/28 13:52
- Last Updated:2021/05/20 09:02
- Category: Energy Dispersive X-ray Fluorescence Spectroscopy(EDX)
-
(EDX) The quantitative values of PbLα and PbLβ1 are signi...
Look at the profile, and check the peaks. •Confirm the PbLα(10.55keV) and PbLβ1(12.62keV) peak intensity ratio. → PbLα : PbLβ1 ≒ 1 : 1 (normal) •If PbLα ... Show Detail
- No:5701
- Date of Issue:2021/04/28 13:50
- Last Updated:2021/05/20 08:59
- Category: Energy Dispersive X-ray Fluorescence Spectroscopy(EDX)
-
(EDX) Negative quantitative value display in quantitative...
At 0ppm, the variance in X-ray intensity will be a negative value if it is below the average intensity. A negative value is treated as a quantitative value ... Show Detail
- No:5699
- Date of Issue:2021/04/28 13:47
- Last Updated:2021/05/20 08:58
- Category: Energy Dispersive X-ray Fluorescence Spectroscopy(EDX)
-
(EDX) Plating film thickness measurement (multiple layers)
• There is no way to specify what layer generated the fluorescent X-rays, and so analysis is not possible if all of the layers are unknown. •If there is 1 u... Show Detail
- No:5697
- Date of Issue:2021/04/28 13:46
- Last Updated:2021/05/20 08:57
- Category: Energy Dispersive X-ray Fluorescence Spectroscopy(EDX)
-
(EDX) Thickness of plating film thickness measurement
The analysis (escape) depth will differ depending on the elements, so it is not easy to specify. As a guideline: •Thick films: Several μm to several doze... Show Detail
- No:5695
- Date of Issue:2021/04/28 13:44
- Last Updated:2021/05/20 08:56
- Category: Energy Dispersive X-ray Fluorescence Spectroscopy(EDX)
-
(EDX) Peaks are detected, but only dotted lines do not id...
• The energy axis is entirely off the profiles. • Calibrate the instrument. Show Detail
- No:5693
- Date of Issue:2021/04/28 13:42
- Last Updated:2021/05/20 08:55
- Category: Energy Dispersive X-ray Fluorescence Spectroscopy(EDX)
-
(EDX) In standard sample analysis, there was a large devi...
Do you perform drift correction? If you are using a calibration curve from a few years ago, it is better to revise the calibration curve. Show Detail
- No:5691
- Date of Issue:2021/04/28 13:41
- Last Updated:2021/05/20 08:54
- Category: Energy Dispersive X-ray Fluorescence Spectroscopy(EDX)
61-70 of Total 132