(PDA) The spectrometer inner pressure has problem
« Cause» • The O-ring of thecondenser lens is notcorrectly installed. • The oil level in thevacuum pump is not thespecified one. • The pump oil in th... Show Detail
- No:6532
- Date of Issue:2021/07/06 10:44
- Last Updated:2021/08/02 16:24
- Category: Optical Emission Spectrometer (PDA)
-
1. Open the [Instrument Status] window. 2. Shut down the software. 3. Shut down Windows and turn OFF the PC and peripheral devices. 4. After tu... Show Detail
- No:6530
- Date of Issue:2021/06/28 14:54
- Category: Optical Emission Spectrometer (PDA)
- Brached FAQ
-
(PDA) Startup after a long period shut down
1. Confirm that the room temperature is in the range of 15 °C to 30 °C. 2. Confirm that the argon gas supply pressure is in the range of140 kPa to 200 kPa. ... Show Detail
- No:6529
- Date of Issue:2021/06/28 14:42
- Category: Optical Emission Spectrometer (PDA)
- Brached FAQ
-
(PDA) Daily Startup / Shutdown
1. Turn ON the power to the printer and monitor. 2. Confirm that the argon gas supply pressure is in the range of140 kPa to 200 kPa. Confirm that the ... Show Detail
- No:6528
- Date of Issue:2021/06/28 12:54
- Category: Optical Emission Spectrometer (PDA)
- Brached FAQ
-
(PDA) Emergency stop and recovering (PDA-7000)
CAUTION: If any problem should occur on the instrument, stop the instrumentby emergency stop and immediately remove the cause of theproblem. 1. Emerge... Show Detail
- No:6527
- Date of Issue:2021/06/28 11:24
- Last Updated:2021/08/02 15:38
- Category: Optical Emission Spectrometer (PDA)
-
(PDA) Replacing the Recalibration Sample
1. Print out the [Analytical Group Information] –[Recalibration] windows for the analytical group tobe changed. 2.Clean the emission table. 3.C... Show Detail
- No:6526
- Date of Issue:2021/06/28 11:01
- Category: Optical Emission Spectrometer (PDA)
- Brached FAQ
-
(PDA) Replacing the Argon Gas Cylinder
When the displayed value for the regulator (1), whichis attached to the argon gas cylinder (hereinafter, thecylinder), becomes 2 MPa or less, replace the cyl... Show Detail
- No:6525
- Date of Issue:2021/06/28 10:40
- Category: Optical Emission Spectrometer (PDA)
-
• Making a backup copy This section describes how to make a backup copy of theanalysis information and results to removable media. 1.Open the [File Mainte... Show Detail
- No:6524
- Date of Issue:2021/06/28 10:10
- Category: Optical Emission Spectrometer (PDA)
- Brached FAQ
-
Lamp test is a way to check behavior of the read-outsystem. Measure the intensities using the lamp test tocheck the stabilization of the whole read-out syste... Show Detail
- No:6519
- Date of Issue:2021/06/25 14:04
- Category: Optical Emission Spectrometer (PDA)
- Brached FAQ
-
(PDA) Manual scanning (Check of entrance slit)
Changes in the environment over the course of timecausing temperature changes inside the system maygive rise to deviations in the peak position of thespectru... Show Detail
- No:6515
- Date of Issue:2021/06/25 13:51
- Category: Optical Emission Spectrometer (PDA)
- Brached FAQ
51-60 of Total 137