FAQ - Frequently Asked Questions
- Products Search > Elemental Analysis
Keyword Search
Frequently Referenced FAQ
FAQ within [ Elemental Analysis ]
-
(EDX) Analysis of samples containing many elements that c...
In quantitative calculations utilizing the FP method, if the samples contains elements that cannot be detected with the EDX (C (carbon), H (hydrogen), O (oxy... Show Detail
- No:5667
- Date of Issue:2021/04/28 13:20
- Last Updated:2021/05/20 08:36
- Category: Energy Dispersive X-ray Fluorescence Spectroscopy(EDX)
-
(EDX) Quantitative analysis with AsKα
Yes, this is possible. The analysis lines (Kα, Lα, etc.)used in quantitative calculations are set in the sensitivity coefficients method files (sysmetal, sy... Show Detail
- No:5669
- Date of Issue:2021/04/28 13:21
- Last Updated:2021/05/20 08:37
- Category: Energy Dispersive X-ray Fluorescence Spectroscopy(EDX)
-
(EDX) About collimator diameter
Basically, the collimator diameter should match the sample size when performing measurements. Cases in which the collimator size does not have to match the ... Show Detail
- No:5671
- Date of Issue:2021/04/28 13:23
- Last Updated:2021/05/20 08:39
- Category: Energy Dispersive X-ray Fluorescence Spectroscopy(EDX)
-
(EDX) Film settings and quantitative values
If the sample contains light elements, then there may be problems with the quantitative values. •The energy of the fluorescent X-rays from the light element... Show Detail
- No:5673
- Date of Issue:2021/04/28 13:24
- Last Updated:2021/05/20 08:41
- Category: Energy Dispersive X-ray Fluorescence Spectroscopy(EDX)
-
(EDX) Analysis of "Si" and "P" in "Cu" alloy
(1) Set the atmosphere to vacuumfor this analysis. →This will improve the sensitivity for light elements. (2) Lower the tube voltagefor the analysis. → ... Show Detail
- No:5675
- Date of Issue:2021/04/28 13:26
- Last Updated:2021/05/20 08:42
- Category: Energy Dispersive X-ray Fluorescence Spectroscopy(EDX)
-
(EDX) Cl detection in "Easy" analysis
The ClKα line overlaps with the RhLα line originating from the X-ray tube. As a result, trace quantities of Cl cannot be detected with a easy analysis method... Show Detail
- No:5677
- Date of Issue:2021/04/28 13:28
- Last Updated:2021/05/20 08:44
- Category: Energy Dispersive X-ray Fluorescence Spectroscopy(EDX)
-
(EDX) "Na" analysis in aqueous solution (without He subst...
Use the filter instillation method for the analysis. In the filter instillation method, droplets of a liquid are dripped on to a filter paper. Then they are... Show Detail
- No:5679
- Date of Issue:2021/04/28 13:29
- Last Updated:2021/05/20 08:46
- Category: Energy Dispersive X-ray Fluorescence Spectroscopy(EDX)
-
(EDX) Thickness of powder sample
Use a micro X-cell. If there is enough sample to extend at least 5mm from the bottom, then the thickness will be sufficient. If the thickness is less than ... Show Detail
- No:5681
- Date of Issue:2021/04/28 13:30
- Last Updated:2021/05/20 08:47
- Category: Energy Dispersive X-ray Fluorescence Spectroscopy(EDX)
-
(EDX) Mylar film and "Mg" analysis
If a Mylar film is used, the Mg intensity will be significantly reduced. Either use a polypropylene film or a pressure molded briquette. Show Detail
- No:5683
- Date of Issue:2021/04/28 13:32
- Last Updated:2021/05/20 08:48
- Category: Energy Dispersive X-ray Fluorescence Spectroscopy(EDX)
-
(EDX) Ceramic plate sample analysis in resin film
Analysis cannot proceed as far as the resin on the opposite side. (1) The ceramic is in the middle, so X-ray intensity from the resin on the opposite side w... Show Detail
- No:5685
- Date of Issue:2021/04/28 13:33
- Last Updated:2021/05/20 08:49
- Category: Energy Dispersive X-ray Fluorescence Spectroscopy(EDX)
51-60 of Total 137