Be sure to use a chemically resistant waste tank because approx. 1 mL/min of sample isdischarged from the sample introduction system of the plasma stand afte... Show Detail
- No:6719
- Date of Issue:2021/07/28 09:58
- Category: Multi type(ICPE)
-
(ICPE) Cleaning the Filter for the Mains Water Supply Con...
NOTE: If the filter's integrated coupling (one-touch joint) is opened while the mainswater supply valve is open, the mains water will gush out, so be sure to... Show Detail
- No:6718
- Date of Issue:2021/07/27 16:17
- Category: Multi type(ICPE)
-
(ICPE) Checking Operation of Safety Devices
The operation of the safety devices used with the ICPE-9800 can be checked as describedbelow. Check the operation of the safety devices on a regular basis (e... Show Detail
- No:6717
- Date of Issue:2021/07/27 16:11
- Category: Multi type(ICPE)
-
(PDA) Basic Points to be Checked When a Problem Occurs
Sometimes, problems are caused by a simple cause. If a problem happens, check the followings first. • Are all power switches turned ON? Confirmtha... Show Detail
- No:6615
- Date of Issue:2021/07/14 13:20
- Category: Optical Emission Spectrometer (PDA)
-
(PDA) The buzzer on the temperature indicator sounds
« Cause» • Abnormal temperature ofthe spectrometer. « Measure» PDA-8000: Confirm that the temperature of theanalysis chamber is 15 °C to 30 °C.... Show Detail
- No:6541
- Date of Issue:2021/07/06 11:06
- Category: Optical Emission Spectrometer (PDA)
-
(PDA) Analysis accuracy is poor
« Cause » • The argon gas leaks during analysis. • The electrode is contaminated. « Measure» Set the sample correctly and performreanalysis. ... Show Detail
- No:6539
- Date of Issue:2021/07/06 11:06
- Last Updated:2021/08/02 16:26
- Category: Optical Emission Spectrometer (PDA)
-
(PDA) Temperature of the spectrometer is unstable, or doe...
« Cause » • Abnormal temperature ofthe spectrometer • The fan supplying hot airhas a problem. « Measure» PDA-8000: Confirm that the temperat... Show Detail
- No:6536
- Date of Issue:2021/07/06 10:54
- Category: Optical Emission Spectrometer (PDA)
-
(PDA) The spectrometer inner pressure has problem
« Cause» • The O-ring of thecondenser lens is notcorrectly installed. • The oil level in thevacuum pump is not thespecified one. • The pump oil in th... Show Detail
- No:6532
- Date of Issue:2021/07/06 10:44
- Last Updated:2021/08/02 16:24
- Category: Optical Emission Spectrometer (PDA)
-
1. Open the [Instrument Status] window. 2. Shut down the software. 3. Shut down Windows and turn OFF the PC and peripheral devices. 4. After tu... Show Detail
- No:6530
- Date of Issue:2021/06/28 14:54
- Category: Optical Emission Spectrometer (PDA)
- Brached FAQ
-
(PDA) Startup after a long period shut down
1. Confirm that the room temperature is in the range of 15 °C to 30 °C. 2. Confirm that the argon gas supply pressure is in the range of140 kPa to 200 kPa. ... Show Detail
- No:6529
- Date of Issue:2021/06/28 14:42
- Category: Optical Emission Spectrometer (PDA)
- Brached FAQ