(ICPE) Checking Operation of Safety Devices
The operation of the safety devices used with the ICPE-9800 can be checked as describedbelow. Check the operation of the safety devices on a regular basis (e... Show Detail
- No:6717
- Date of Issue:2021/07/27 16:11
- Category: Multi type(ICPE)
-
(PDA) Basic Points to be Checked When a Problem Occurs
Sometimes, problems are caused by a simple cause. If a problem happens, check the followings first. • Are all power switches turned ON? Confirmtha... Show Detail
- No:6615
- Date of Issue:2021/07/14 13:20
- Category: Optical Emission Spectrometer (PDA)
-
(PDA) The buzzer on the temperature indicator sounds
« Cause» • Abnormal temperature ofthe spectrometer. « Measure» PDA-8000: Confirm that the temperature of theanalysis chamber is 15 °C to 30 °C.... Show Detail
- No:6541
- Date of Issue:2021/07/06 11:06
- Category: Optical Emission Spectrometer (PDA)
-
(PDA) Analysis accuracy is poor
« Cause » • The argon gas leaks during analysis. • The electrode is contaminated. « Measure» Set the sample correctly and performreanalysis. ... Show Detail
- No:6539
- Date of Issue:2021/07/06 11:06
- Last Updated:2021/08/02 16:26
- Category: Optical Emission Spectrometer (PDA)
-
(PDA) Temperature of the spectrometer is unstable, or doe...
« Cause » • Abnormal temperature ofthe spectrometer • The fan supplying hot airhas a problem. « Measure» PDA-8000: Confirm that the temperat... Show Detail
- No:6536
- Date of Issue:2021/07/06 10:54
- Category: Optical Emission Spectrometer (PDA)
-
(PDA) The spectrometer inner pressure has problem
« Cause» • The O-ring of thecondenser lens is notcorrectly installed. • The oil level in thevacuum pump is not thespecified one. • The pump oil in th... Show Detail
- No:6532
- Date of Issue:2021/07/06 10:44
- Last Updated:2021/08/02 16:24
- Category: Optical Emission Spectrometer (PDA)
-
1. Open the [Instrument Status] window. 2. Shut down the software. 3. Shut down Windows and turn OFF the PC and peripheral devices. 4. After tu... Show Detail
- No:6530
- Date of Issue:2021/06/28 14:54
- Category: Optical Emission Spectrometer (PDA)
- Brached FAQ
-
(PDA) Startup after a long period shut down
1. Confirm that the room temperature is in the range of 15 °C to 30 °C. 2. Confirm that the argon gas supply pressure is in the range of140 kPa to 200 kPa. ... Show Detail
- No:6529
- Date of Issue:2021/06/28 14:42
- Category: Optical Emission Spectrometer (PDA)
- Brached FAQ
-
(PDA) Daily Startup / Shutdown
1. Turn ON the power to the printer and monitor. 2. Confirm that the argon gas supply pressure is in the range of140 kPa to 200 kPa. Confirm that the ... Show Detail
- No:6528
- Date of Issue:2021/06/28 12:54
- Category: Optical Emission Spectrometer (PDA)
- Brached FAQ
-
(PDA) Emergency stop and recovering (PDA-7000)
CAUTION: If any problem should occur on the instrument, stop the instrumentby emergency stop and immediately remove the cause of theproblem. 1. Emerge... Show Detail
- No:6527
- Date of Issue:2021/06/28 11:24
- Last Updated:2021/08/02 15:38
- Category: Optical Emission Spectrometer (PDA)
41-50 of Total 132