Check the tank periodically to make sure the following. ・ The waste liquid is not overflow. ・ The drain tube does not come in contact with the liquid surfa... Show Detail
- No:6772
- Date of Issue:2021/08/18 15:08
- Last Updated:2021/09/13 09:15
- Category: Sequential type(ICPS)
-
Approximately 10L of pure water is used as the coil cooling water. Replace the water withnew one when fur forms in it. WARNIGN:You may get an electric... Show Detail
- No:6771
- Date of Issue:2021/08/18 15:08
- Category: Sequential type(ICPS)
-
(ICPS) Cleaning of Orifice for Axial Observation(ICPS-7510)
NOTE: Cleaning the orifice prevents contamination of a sample and abnormaldischarge. WARNING:Never touch the inside of the stand for three minutes after e... Show Detail
- No:6769
- Date of Issue:2021/08/18 15:05
- Last Updated:2022/01/17 15:47
- Category: Sequential type(ICPS)
-
(ICPS) Maintenance in the plasma stand(ICPS-7510)
1.Cleaning part (1) Sheet copper connected with RF coil and the connecting part (2) Connecting part of this sheet copper and matching-Box.And the circumf... Show Detail
- No:6770
- Date of Issue:2021/08/18 15:04
- Category: Sequential type(ICPS)
-
(ICPS) Cleaning of Folding Mirror for Axial Observation(I...
WARNING:You may get a burn. Before starting the procedures below, turn OFF theplasma and wait until the water cooled jacket is cooled sufficiently (at least3... Show Detail
- No:6768
- Date of Issue:2021/08/18 14:50
- Category: Sequential type(ICPS)
-
(ICPS) Operation Check of Safety Device
WARNING: You may a burn. Don't touch the exhaust pipe on a plasma stand, or don't look in. 1) Argon gas pressure switch If you stop supplying argo... Show Detail
- No:6766
- Date of Issue:2021/08/18 12:48
- Category: Sequential type(ICPS)
- Brached FAQ
-
(ICPE) Filter Cleaning and Replacement (Autosampler)
If the filter is clogged, remove the cover on the left side of the ASC-9800 and clean the filter. If the filter has deteriorated, replace it with a new filt... Show Detail
- No:6742
- Date of Issue:2021/08/03 10:56
- Category: Multi type(ICPE)
- Brached FAQ
-
(ICPE) Replacing the Peristaltic Pump Tubing (Autosampler)
If the flow rate of the peristaltic pump used to supply rinse solution decreases, replace the tubing in the peristaltic pump according to the following proce... Show Detail
- No:6741
- Date of Issue:2021/08/03 10:51
- Category: Multi type(ICPE)
- Brached FAQ
-
(ICPE) Mounting the Nozzle (Autosampler)
1. Insert the nozzle (Parts number: 206-50290-91) into the nozzle hole located on the arm. 2. Lower the nozzle tip by 103 mm from the bottom of the arm an... Show Detail
- No:6740
- Date of Issue:2021/08/03 10:21
- Category: Multi type(ICPE)
- Brached FAQ
-
(ICPE) The spectral intensity for elements is high even w...
• The blank liquid is dirty. • The torch is dirty. • The chamber is dirty. Click here for "Replacing and Inspecting Glassware, Sample Introduction Tub... Show Detail
- No:6739
- Date of Issue:2021/07/30 10:11
- Category: Multi type(ICPE)