(ICPS) Element spectrum intensity is high although blank
・Contamination of blank liquid ・Contamination of the torch ・Contamination of the chamber ・Rinsing time is short. Click here for "Check and Replacemen... Show Detail
- No:6785
- Date of Issue:2021/08/19 09:43
- Category: Sequential type(ICPS)
-
(ICPS) E1, E2, E3 error (Auto sampler: AS-9)
E1: A moving erroe of X-axis E2: A moving erroe of Y-axis E3: A moving erroe of Z-axis If above error indication appeared, it is a movement erro... Show Detail
- No:6787
- Date of Issue:2021/08/19 10:14
- Category: Sequential type(ICPS)
-
Please refer to the YouTube URL below. Mounting of the Glassware Parts: https://youtu.be/f-3BMN09U4M Detachment of the Glassware Parts: https://youtu.be/ny... Show Detail
- No:11596
- Date of Issue:2025/01/30 09:29
- Last Updated:2025/01/30 09:32
- Category: ICP Mass Spectrometer (ICP-MS)
-
(ICPE) About internal standard (IS)
Non-spectral interferences may occur when a sample has a high matrix concentration. These include physical interferences and ionization interferences. Physic... Show Detail
- No:12257
- Date of Issue:2026/05/08 13:42
- Category: Multi type(ICPE)