FAQ - Frequently Asked Questions
- Products Search > Elemental Analysis
Keyword Search
Frequently Referenced FAQ
FAQ within [ Elemental Analysis ]
-
(ICPS) Element spectrum intensity is high although blank
・Contamination of blank liquid ・Contamination of the torch ・Contamination of the chamber ・Rinsing time is short. Click here for "Check and Replacemen... Show Detail
- No:6785
- Date of Issue:2021/08/19 09:43
- Category: Sequential type(ICPS)
-
(ICPS) E1, E2, E3 error (Auto sampler: AS-9)
E1: A moving erroe of X-axis E2: A moving erroe of Y-axis E3: A moving erroe of Z-axis If above error indication appeared, it is a movement erro... Show Detail
- No:6787
- Date of Issue:2021/08/19 10:14
- Category: Sequential type(ICPS)
-
Please refer to the YouTube URL below. Mounting of the Glassware Parts: https://youtu.be/f-3BMN09U4M Detachment of the Glassware Parts: https://youtu.be/ny... Show Detail
- No:11596
- Date of Issue:2025/01/30 09:29
- Last Updated:2025/01/30 09:32
- Category: ICP Mass Spectrometer (ICP-MS)
131-133 of Total 133