(AA) GFA, Cleaning and Replacing Procedure for Graphite C...
1. Press the eject arm to release the lock and remove the graphite tube. 2. Pull the window plate sockets out from the cooling blocks. 3. Wipe off soot dep... Show Detail
- No:5434
- Date of Issue:2021/03/31 11:28
- Last Updated:2022/07/08 15:31
- Category: Atomic Absorption Spectroscopy(AA)
- Brached FAQ
-
(AA) GFA, Cleaning the Window Plate Sockets and Window Pl...
WARNING: • RISK OF ELECTRIC SHOCK:Before attaching or detaching the graphite furnace or heating cables, make sure that the power is off. • The window pla... Show Detail
- No:5555
- Date of Issue:2021/03/31 11:10
- Last Updated:2022/07/08 15:37
- Category: Atomic Absorption Spectroscopy(AA)
-
(AA) GFA, Mounting the Graphite Cap and Holder
1. Mount the graphite cap. 2. Press the release lever to release the lock of the furnace and open the right-hand cooling block fully to theright. 3. Insert... Show Detail
- No:5430
- Date of Issue:2021/03/31 10:51
- Last Updated:2022/07/08 15:20
- Category: Atomic Absorption Spectroscopy(AA)
-
Cause1: Deposits have become attached tothe slot in the burner head. Corrective Action1: Wash the burner head.Click here forBurner Maintenance Caus... Show Detail
- No:5315
- Date of Issue:2021/03/26 13:27
- Last Updated:2022/07/07 11:14
- Category: Atomic Absorption Spectroscopy(AA)
-
(AA) The absorbance or energy value has fallen very consi...
Cause1:Either the hollow cathode lamp for thetarget element has not been installed,or it has been installed in a socketthat differs from the analysisconditio... Show Detail
- No:5314
- Date of Issue:2021/03/26 13:25
- Last Updated:2022/07/07 11:13
- Category: Atomic Absorption Spectroscopy(AA)
-
Cause: The hollow cathode lamp does notlight stably. Corrective Action:After performing a line search, leavethe lamp for about 15 minutes until itwarms up... Show Detail
- No:5313
- Date of Issue:2021/03/26 12:04
- Last Updated:2022/07/07 11:12
- Category: Atomic Absorption Spectroscopy(AA)
-
(AA) The absorbance is -1(minus one)
Cause1: The light paths of the hollow cathodelamp and the deuterium lamp do notcoincide.(When the BGC-D2 mode is set) Corrective Action1:Adjust the positi... Show Detail
- No:5312
- Date of Issue:2021/03/26 12:02
- Last Updated:2022/07/07 11:11
- Category: Atomic Absorption Spectroscopy(AA)
-
(AA) The line search / beam balancing failed
Cause1: There is something blocking the lightpath in the atomizer. Corrective Action1: If there is something blocking the lightpath, remove it. Cause... Show Detail
- No:5306
- Date of Issue:2021/03/26 12:00
- Last Updated:2022/07/07 11:10
- Category: Atomic Absorption Spectroscopy(AA)
-
(AA) The signal-to-noise ratio is bad
Cause1:Either the hollow cathode lamp for thetarget element has not been installed,or it has been installed in a socketthat differs from the analysisconditio... Show Detail
- No:5268
- Date of Issue:2021/03/26 11:54
- Last Updated:2022/07/07 11:09
- Category: Atomic Absorption Spectroscopy(AA)
-
(AA) A wavelength origin error occurs at initialization
Cause1: There is something blocking the lightpath in the atomizer. Corrective Action1: If there is something blocking the lightpath, remove it. Cau... Show Detail
- No:5267
- Date of Issue:2021/03/26 11:42
- Last Updated:2022/07/07 11:08
- Category: Atomic Absorption Spectroscopy(AA)