FAQ within [ Elemental Analysis ]
-
(PDA) Basic Points to be Checked When a Problem Occurs
Sometimes, problems are caused by a simple cause. If a problem happens, check the followings first. • Are all power switches turned ON? Confirmtha... Show Detail
- No:6615
- Date of Issue:2021/07/14 13:20
- Category: Optical Emission Spectrometer (PDA)
-
(ICPE) Checking Operation of Safety Devices
The operation of the safety devices used with the ICPE-9800 can be checked as describedbelow. Check the operation of the safety devices on a regular basis (e... Show Detail
- No:6717
- Date of Issue:2021/07/27 16:11
- Category: Multi type(ICPE)
-
(ICPE) Cleaning the Filter for the Mains Water Supply Con...
NOTE: If the filter's integrated coupling (one-touch joint) is opened while the mainswater supply valve is open, the mains water will gush out, so be sure to... Show Detail
- No:6718
- Date of Issue:2021/07/27 16:17
- Category: Multi type(ICPE)
-
Be sure to use a chemically resistant waste tank because approx. 1 mL/min of sample isdischarged from the sample introduction system of the plasma stand afte... Show Detail
- No:6719
- Date of Issue:2021/07/28 09:58
- Category: Multi type(ICPE)
-
(ICPE) Cleaning the Fan Filters
The fan filter on the left side of the main unit must be cleaned on a regular basis. Clean this filter once every three months. Parts Number: 211-88556 F... Show Detail
- No:6720
- Date of Issue:2021/07/28 10:19
- Last Updated:2021/10/04 09:49
- Category: Multi type(ICPE)
- Brached FAQ
-
(ICPE) Maintenance for Vacuum Pump
The molecular sieve in the vacuum pump at the back of the ICPE-9800 can be replacedwhile maintaining a vacuum state in the spectrometer. Replace the molecul... Show Detail
- No:6721
- Date of Issue:2021/07/28 10:55
- Last Updated:2021/10/04 09:49
- Category: Multi type(ICPE)
- Brached FAQ
-
(ICPS) Maintenance of Vacuum Pump
The molecular sieve can be replaced by a new one with the spectrometerkept at a vacuum status. Parts number: 239-07003-08 Molecular siev Parts numb... Show Detail
- No:6775
- Date of Issue:2021/07/28 10:55
- Category: Sequential type(ICPS)
- Brached FAQ
-
(ICPE) Replacing and Inspecting Glassware, Sample Introdu...
WARNING: There is a risk of burn injury. Do NOT touch the plasma standinterior for 3 minutes after the plasma is extinguished. CAUTION: Glass parts break ... Show Detail
- No:6724
- Date of Issue:2021/07/29 14:03
- Category: Multi type(ICPE)
- Brached FAQ
-
(ICPS) Check and Replacement of Glassware
1) Holding the chamber, remove the ball joint clip. 2) Loosen the clamp of torch stand, pull out the torch downward and remove the tube connec... Show Detail
- No:6767
- Date of Issue:2021/07/29 14:03
- Category: Sequential type(ICPS)
- Brached FAQ
-
(ICPE) Replacing the Peristaltic Pump Tube
Since the sample pressure feed tube is constrained by the rotor and guide, it is subjectto plastic deformation and eventually samples will no longer feed thr... Show Detail
- No:6726
- Date of Issue:2021/07/29 14:31
- Category: Multi type(ICPE)
- Brached FAQ