(EDX) The instrument is not turned ON.
Check Points1:Check that the power cable isconnected correctly. Remedies1:Connect the power cable correctly. Check Points2:Check that power is su... Show Detail
- No:5615
- Date of Issue:2021/04/07 10:28
- Category: Energy Dispersive X-ray Fluorescence Spectroscopy(EDX)
-
(EDX) Changing and Supplying Oil to the Rotary Pump (When...
The rotary pump used as an auxiliary pump for the evacuation system requires oil change every one and a halfyears*. Without oil change, insufficient vacuum, ... Show Detail
- No:5614
- Date of Issue:2021/04/07 10:21
- Category: Energy Dispersive X-ray Fluorescence Spectroscopy(EDX)
-
1. X-rays will be generated. Close the sample chamber cover. 2. Double click theor icon on the Desktop to launch the software. 3. Initialize th... Show Detail
- No:5612
- Date of Issue:2021/04/06 14:38
- Category: Energy Dispersive X-ray Fluorescence Spectroscopy(EDX)
- Brached FAQ
-
(AA) ASC, Cleaning the Syringe
The syringe may be dismounted for cleaning of its inside wall using the following procedure. 1. Remove the rinse liquid intake tube from the rinse liqui... Show Detail
- No:5429
- Date of Issue:2021/04/05 09:28
- Last Updated:2022/07/07 13:45
- Category: Atomic Absorption Spectroscopy(AA)
-
(AA) ASC, Rinsing the ASC Nozzle Flow Path
Contamination of the inside wall of the PTFE tube, the inside wall of the syringe, the plunger tip and liquidcontact surfaces of the solenoid valve are likel... Show Detail
- No:5426
- Date of Issue:2021/04/05 09:25
- Last Updated:2022/07/07 13:43
- Category: Atomic Absorption Spectroscopy(AA)
-
(AA) ASC, Replacing PTFE Tube for Nozzle ASSY, G (Optional)
Parts Number:016-37551PTFE tube, 0.60×0.25 When the nozzle ASSY, G for furnace measurement has been used for a long time, the tip of the PTFE tube for... Show Detail
- No:5423
- Date of Issue:2021/04/05 09:24
- Last Updated:2022/07/07 13:42
- Category: Atomic Absorption Spectroscopy(AA)
-
(AA) ASC, Replacement of Peristaltic Pump Tube
Parts Number:042-00405-24Pump tube Parts Number:042-00405-11Pump head When the flow rate of peristaltic pump is decreased, replace the pump tube wi... Show Detail
- No:5420
- Date of Issue:2021/04/05 09:17
- Last Updated:2022/07/07 13:39
- Category: Atomic Absorption Spectroscopy(AA)
-
Parts Number:046-00359-01 Pipet tip When the pipet tip for furnace / flame micro sampling measurement has been used for a long time, the pipet tipand inne... Show Detail
- No:5419
- Date of Issue:2021/04/05 09:12
- Last Updated:2023/12/27 15:41
- Category: Atomic Absorption Spectroscopy(AA)
-
Parts Number:208-97211-01 Syringe (Capacity 250 μL) Parts Number:046-00043-15 Tip, TEF025 Select [Instrument] - [Maintenance] - [ASC Maintenanc... Show Detail
- No:5416
- Date of Issue:2021/03/31 14:42
- Last Updated:2022/07/08 09:08
- Category: Atomic Absorption Spectroscopy(AA)
-
(AA) GFA, Replacing the Graphite Tube
WARNING:Immediately after heating the graphite tube is hot, so leave it for at least 3 minutes to cool beforeremoving it. If you touch the graphite tube imm... Show Detail
- No:5431
- Date of Issue:2021/03/31 13:50
- Category: Atomic Absorption Spectroscopy(AA)
- Brached FAQ