Flat panel detector calibration is a function used to reduce unevenness and noise in the image.
Perform calibration if unevenness or noise in the image becomes noticeable during instrument use.
The duration of the flat panel detector calibration ([Advanced]) is approximately 5 minutes.
NOTE: Use this function in administrator mode
Click here for "(SMX-1010/1020) Switching to Administrator Mode".
1. Display the [Detector Calibration] window.
1) Click the [Manage] tab.
2) Click [Detector Calibration].
2. Select the detector calibration type.
1) Click [Simple] or [Advanced].
• [Simple]: Performs flat panel detector calibration at the set tube voltage.
• [Advanced]: Performs flat panel detector calibration at all selectable tube voltages.
2) Click [Start].
Flat panel detector calibration starts.
3. When flat panel detector calibration is complete, click [OK].
If necessary, observe the sample and check that unevenness and noise in the image have disappeared.