(SMX-1010/1020) Outputting the Instrument Information
Output the information related to the usage status of the instrument. If a problem occurs with the instrument, you may be required to output the instrumenti... Show Detail
- No:8634
- Date of Issue:2022/05/17 09:26
-
(SMX-1010/1020) The X-RAY ON/OFF button remains gray and ...
If the condition does notimprove even after performing Action 1 -3,contacta Shimadzu sales or service representative. Cause1: Is the front door comple... Show Detail
- No:8628
- Date of Issue:2022/04/28 14:32
-
(SMX-1010/1020) Exterior Image trouble
Trouble1: No exterior image isdisplayed. Cause1-1: The exterior image cameramay be faulty. Cause1-2: There may be a faultyconnection or disconnectioninsi... Show Detail
- No:8626
- Date of Issue:2022/04/28 14:14
-
(SMX-1010/1020) Cross-section images are blurry
Cause1:Is the sample moving duringimage scanning? Action1:Fix the sample in place withtape or other means. Cause2: Cross-section images can beblurr... Show Detail
- No:8625
- Date of Issue:2022/04/28 14:10
-
(SMX-1010/1020) There is unevenness or noise in X-ray images
Cause: Unevenness and noise influoroscopic images may becaused by deterioration of theflat panel detector over time. Action:Calibrate the flat paneldetect... Show Detail
- No:8624
- Date of Issue:2022/04/28 14:05
-
(SMX-1010/1020) The X-ray image is completely black
Cause1: Are X-rays emitted? Action1: Emit X-rays. Cause2:The X-ray image can becompletely black if X-rayoutput is too low. Action2: Change the tub... Show Detail
- No:8623
- Date of Issue:2022/04/28 14:03
-
(SMX-1010/1020) The stage does not move
Cause1:Is the SYSTEM key switch ON? Action1: Turn ON the SYSTEM keyswitch. Cause2:Is the OPERATE switch ON? Action2:Turn ON the OPERATE switch. ... Show Detail
- No:8622
- Date of Issue:2022/04/28 14:01
-
(SMX-1010/1020) The interior light does not light up
Cause1:Is the SYSTEM key switch ON? Action1: Turn ON the SYSTEM keyswitch. Cause2:Is the OPERATE switch ON? Action2: Turn ON the OPERATE switc... Show Detail
- No:8621
- Date of Issue:2022/04/28 13:55
-
(SMX-1010/1020) Performing the Pre-Operation Check and In...
In the pre-operation check and initialization process, you will check the front door, returnthe stage to its origin, and perform the aging of the X-ray gener... Show Detail
- No:8617
- Date of Issue:2022/04/28 13:34
-
(SMX-1010/1020) Performing Flat Panel Detector Calibration
Flat panel detector calibration is a function used to reduce unevenness and noise in theimage. Perform calibration if unevenness or noise in the image becom... Show Detail
- No:8620
- Date of Issue:2022/04/28 10:30