• No : 8598
  • Date of Issue : 2022/04/28 09:20
  • Print

(SMX-6010) Adjusting the Filament Value

(SMX-6010) Adjusting the Filament Value
Category : 

Answer

Filament value adjustment is the operation in which the current that flows through the filament is adjusted to an optimal value in order to enable the filament to be used over a long period of time and to obtain stable images.
Although the filament value can be adjusted automatically after aging at instrument startup and aging after filament replacement, it can also be adjusted at any time.
NOTE: Use this function in administrator mode.
  Click here for "(SMX-6010) Switching to Administrator Mode".
 
1. Display the [Filament Value Adjustment] window.
  1) Click the [X-ray Generator] tab.
  2) Click [Filament Value Adjustment].
  
 
2. Click [Run].
  
 
3. Confirm that no sample is set on the stage.
 
4. Click [Start].
  
 
5. When filament value adjustment is complete, click [OK].