(EZ-Test) A servo error occurred during jog operations

(EZ-Test) A servo error occurred during jog operations
Category : 

Answer

• The motor may be overloaded. Confirm there are no obstructions caught between the crosshead and table. Also confirm that nothing is wrapped around the ball screw threads.