(EHF) The set displacement amplitude does not appear

(EHF) The set displacement amplitude does not appear
Energy Saving Power Unit Software Trouble
Windows Software for 4830
Category : 

Answer

Automatic control during testing with Type 1 calculates with the supply pressure set in the detailed supply pressure. If the supply pressure of the hydraulic source during the test is lower than the set value, the expected automatic control cannot be performed. Make sure that the supply pressure of the hydraulic source exceeds the set value.