(EHF) Control switching is not possible
(EHF) Control switching is not possible
EHF-E, U, L Series
Servo Controller 4830
Category :
Products Search
>
Materials Testing & Inspection
>
Servopulser
>
EHF Series
Answer
If you cannot solve this problem by checking possible causes and taking appropriate actions described below,
contact a Shimadzu sales or service representative
.
Cause1: The sensor cable is not connected, or almost disconnected.
Action1: Connect the sensor cable securely.
Cause2: The load cell cable is disconnected, or almost disconnected.
Action2: Connect the cable securely. A warning message is displayed when the servo controller starts.
Cause3: Select “Invalid”.
Action3: Please select "Valid".
Click here
for "(EHF) TD calibration".