In this machine, the past 50 event logs that are generated are saved. Event logs are subdivided intothe following classes.
• Behavior of the testing machine(Start of test, End of test, Return, PC connect/disconnect)
• Generated alarm
• Calibration
• Initialization
1. Touch the [System] menu key on the LCD display. The System screen appears.

2. Touch the [Maintenance] key. The Maintenance screen appears. Select the [Log] tab.
you can switch the page with

or

. Event and initiation time of the event appear in the display.