« Cause »
• Abnormal temperature of the spectrometer
• The fan supplying hot air has a problem.
« Measure »
PDA-8000: Confirm that the temperature of the analysis chamber is 15 °C to 30 °C. After that, shut down the system, and then restart it.
•
Click here for Startup after a long period shut down
PDA-7000: Check that the room temperature is within 10 - 28°C and reset the AIRCON switch on the switch panel.