• No : 6059
  • Date of Issue : 2021/05/18 15:33
  • Last Updated : 2021/06/16 09:19
  • Print

AOC-20s expansion and control error[Error Code:6015]

The following causes are possible.
• The extension motor has a problem.
• A vial with dimensions different from those of the vial used for teaching is used.
Category : 

Answer

Redo the teaching using the vial to be used for actual analysis.
   [In case of AOC-30i]
      AOC-30i_AOC-20sU_SamplerTeaching.pdf
   [In case of AOC-20i]
      AOC-20i_AOC-20s_SamplerTeaching.pdf