• No : 5999
  • Date of Issue : 2021/05/18 15:36
  • Last Updated : 2023/12/13 13:23
  • Print

CARm calc. ratio out of range[Error Code:5075 5076 5108 5109]

The carrier gas total flow calculated from split ratio program, etc. is outside the set range.
Category : 

Answer

Change the split ratio so that the following equation is satisfied. 
(Total flow) = (Column flow) + (Column flow * Split ratio) + (Purge flow)