• No : 5931
  • Date of Issue : 2021/05/18 15:36
  • Last Updated : 2021/06/08 11:17
  • Print

Clean up program start error [Error Code:4305]

Clean up program cannot be started because oven temperature is not ready.
Category : 

Answer

Set the initial temperature of the column oven at higher values than room temperature by more than 2℃.
(Example, Set the temperature at more than 27 ℃ when the room temperature is 25℃)
 
If the error appears repeatedly even after checking the above, turn off the main power of the instrument, stop gas supply,
and contact your Shimadzu sales/service representative.