• No : 4965
  • Date of Issue : 2021/02/05 10:54
  • Print

(HPLC) Check the Condition of Consumables (System Check)

(HPLC) Check the Condition of Consumables (System Check)
Category : 

Answer

By executing the system check, you can check the usage of consumables for the entire system.
 
• LCsolution, LabSolutions
To open the [System Check] sub-window, click the  (System Check) icon on the [Main] assistant bar of the [Realtime Analysis] program, or click  (System Check) on the [Instrument Control] toolbar.
Or, from [Instruments] → [System Check] to enter the [System Check] winodw.
HINT:
LCsolution: When the check level is set to [Complete], the lamp energy and wavelength check of the detector is performed. In the case of [Normal], the result of the previous execution is displayed.
LabSolutions: If you check [Advanced] → [UV Detector Wavelength Check], the lamp energy and wavelength check of the detector will be executed. If unchecked, the result of the previous execution is displayed.
 
• Waters Empower
To open the [System Check] window, select [System Check] from the menu displayed by clicking  (Control) button in the Shimadzu LC Control Panel.
HINT: When the check level is set to [Complete], the lamp energy and wavelength check of the detector is performed. In the case of [Normal], the result of the previous execution is displayed.
 
• For non-workstations
CBM-20A(lite), CBM-40: You can perform a system check from the Web controller also.
SCL-40, i-Series(LC-2030/40/50/60): You can execute the system check from the screen operation also.
 
*Waters® and Empower® Software are registered trademarks of Waters in the United States and other countries.