(HPLC) RID "OVER" is displayed

(HPLC) RID "OVER" is displayed
Category : 

Answer

• "OVER" is displayed in the "range" field of the initial screen.
  Cause: The recorder output has exceeded the scale limit.
  Actions: Press [zero].
 
• "OVER" is displayed in the "balance" field of the initial screen.
  Cause: Optical balance is shifted.
  Actions: Press [bal] to balance the optical system.
    * "OVER" may be displayed when peak of chromatogram is detected.
      This is not a problem. In this case, you need not balance the optical system.