(GCMS) "Detector saturated" message appears during analysis

(GCMS) "Detector saturated" message appears during analysis.
Category : 

Answer

1. Turn on the filament after solvent has completely eluted. If peaks elute before the solvent, use a time program to turn the filament off and on.
2. Decrease the detector voltage.
3. Reduce the sample concentration by making a dilution or reduce the injection volume.
4. If unavoidable, switch the filament OFF.