The status remains [Pause] and the analysis does not advance to the next step

The status remains [Pause] and the analysis does not advance to the next step.
Category : 

Answer

«Main Cause»
The [Event In 1] setting is [Ready In]. 
 
«Countermeasure»
Reconsider the [Event In 1] setting in system controller [Environment setting] settings.