Instrument is not recognized when connected via an optical link cable to an SCL-40 unit not connected to workstation software

Instrument is not recognized when connected via an optical link cable to an SCL-40 unit not connected to workstation software.
Category : 

Answer

«Main Cause»
Link mode setting is [Fixed]
(tends to occur if workstation software was force-quit). 
 
«Countermeasure»
Set link mode to [Auto].