Products Search
>
HPLC
>
Detector
>
Error Code:[0x5A7A]
Back
No : 4109
Date of Issue : 2020/12/02 16:46
Print
Error Code:[0x5A7A]
Error Message: PDA detector: The temperature sensor in the lamp housing did not operate normally, so the cell temperature control was stopped.
Category :
Products Search
>
HPLC
>
Detector
Answer
«Main Cause»
A broken circuit may have occurred in the lamp housing sensor.
«Countermeasure»
Switch the instrument power OFF and contact a
Shimadzu sales or service representative
.