Products Search
>
HPLC
>
Column Oven
>
Error Code:[0x5403] ERROR EXTERNAL INPUT
Back
No : 4025
Date of Issue : 2020/12/02 16:45
Print
Error Code:[0x5403] ERROR EXTERNAL INPUT
Error Message: Column oven: The temperature control was stopped because an error signal was inputted to external input terminal.
Category :
Products Search
>
HPLC
>
Column Oven
Answer
«Main Cause»
Temperature control stopped because an error signal was input via the column oven external input terminal.
«Countermeasure»
Check the connected external device and remove the cause of the error.