Products Search
>
HPLC
>
Autosampler
>
Error Code:[0x5257] ERROR PFILE WAIT.RT RNG
Back
No : 3968
Date of Issue : 2020/12/02 16:44
Print
Error Code:[0x5257] ERROR PFILE WAIT.RT RNG
Error Message: Autosampler: The wait time in the pretreatment program is out of range. Change the wait time to within the range.
Category :
Products Search
>
HPLC
>
Autosampler
Answer
«Main Cause»
The wait time specified by the pretreatment program exceeds the setting range.
«Countermeasure»
Change the setting to a value within the specified range.