Products Search
>
HPLC
>
Autosampler
>
Error Code:[0x5252] ERROR PFILE GOTO FOR.B
Back
No : 3963
Date of Issue : 2020/12/02 16:44
Print
Error Code:[0x5252] ERROR PFILE GOTO FOR.B
Error Message: Autosampler: A line number which is between the FOR command and NEXT command is specified with the GOTO command. Correct the line number of the GOTO command.
Category :
Products Search
>
HPLC
>
Autosampler
Answer
«Main Cause»
The GOTO command in the pretreatment program specifies a line number that is within a FOR-NEXT range.
«Countermeasure»
Revise the line number for the GOTO command.