Error Code:[0x5252] ERROR PFILE GOTO FOR.B

Error Message: Autosampler: A line number which is between the FOR command and NEXT command is specified with the GOTO command. Correct the line number of the GOTO command.
Category : 

Answer

«Main Cause»
The GOTO command in the pretreatment program specifies a line number that is within a FOR-NEXT range.
 
«Countermeasure»
Revise the line number for the GOTO command.