Error Code:[0x910A] FATAL LEAK SENSOR
Error Message: Pump: The pumping was stopped because the leak sensor cannot be recognized correctly.
Category :
Products Search
>
HPLC
>
Solvent Delivery Module
Answer
«Main Cause»
A leak sensor short circuit may have occurred.
«Countermeasure»
Switch the instrument power OFF and contact a
Shimadzu sales or service representative
.