Error Code:[0x5F04] FATAL USER DATA
Error Message: The saved user data was initialized because the saved data was not correct. There is no effect on instrument performance, but check the parameter.
Category :
Products Search
>
HPLC
>
Common
Answer
«Main Cause»
Data was initialized because the saved user data was incorrect.
«Countermeasure»
Instrument performance is not affected, but double-check respective parameter settings.