«Main Cause»
A seal problem occurred in the degassing unit vacuum flow channels.
«Countermeasure»
Check the vacuum pressure in the degassing unit. If the vacuum pressure status is [NG], switch the instrument power OFF and then switch it ON again. If the vacuum pressure status is [OK], the degassing unit was automatically restored and there is no problem.