Error Code:[0x80F3]

Error Message: PDA detector's IP Address has been changed and the connection between system controller and PDA detector has been disconnected. Check the communication Settings (parent-child relationship).
Category : 

Answer

«Main Cause»
The IP address for the PDA detector was changed during use.
 
«Countermeasure»
The parent-child relationship might have failed between the system controller and PDA detector. Check the parent-child relationship.